Reproducibility vs Repeatability: Distinguishing Between Two Often Confused Types of Precision

MyQCLabOctober 4, 202637 views
Reproducibility vs Repeatability: Distinguishing Between Two Often Confused Types of Precision

Understand the differences between repeatability and reproducibility in laboratory precision, complete with a nested ANOVA calculator to distinguish between them.

Category: Advanced Quality Instruments | Estimated reading time: 10 minutes

The "Hidden" Precision Behind CV

In the Preliminary Test article, the CV is calculated from 30 control material measurements under consistent normal conditions — the same instrument, the same operator, and days that are (relatively) close together. That CV figure is valid and important, but it only tells one side of true precision: how consistent the instrument is when everything is uniform.

The unanswered question: what if the conditions are not uniform? What if measurements are taken on different days, with reasonable fluctuations in room temperature, by operators on rotating shifts? Is the precision just as good? This is where Repeatability and Reproducibility — two concepts often mistakenly used interchangeably — must be understood as two distinct things.

Definitions

Repeatability is the ability of an instrument or examiner to produce consistent results when the test is repeated under the same conditions — the same instrument, location, operator, and time (or close proximity).

Reproducibility is the ability to produce consistent results when the test is performed under different conditions — different operators, different instruments, different locations, or different times (e.g., between days).

Both are part of the precision evaluation in laboratory testing methods — the difference lies in how many sources of variation are included in the calculation.

Brief History

This concept stems from the analytical method validation principles widely used in the pharmaceutical industry, clinical laboratories, and scientific research. Its standardization was strengthened by the international body ISO (International Organization for Standardization), specifically in the ISO 5725 document series (Accuracy — trueness and precision — of Measurement Methods and Results).

It is important to understand that ISO 5725 is not a single document, but a multi-part series: ISO 5725-1 establishes general principles and basic definitions (including the formal definitions of Repeatability and Reproducibility used in this article), while ISO 5725-2 contains the basic methods for determining repeatability and reproducibility standard deviations through collaborative interlaboratory experiments. In the specific context of clinical laboratories, similar principles were adopted and adapted by CLSI EP05-A3, a document also referenced in the Preliminary Test article — showing that both topics actually originate from the same statistical framework.

Reasons for Use

  • Ensuring the reliability of laboratory results.
  • Assessing the precision of test methods, both internally (repeatability) and externally (reproducibility).
  • As part of method validation, instrument verification, and laboratory quality assurance processes.

Advantages

  • Provides a quantitative evaluation of test result stability.
  • Allows for the identification of technical variability, whether from the instrument, technician, or environment — and most importantly, isolating the contribution of each source of variation, rather than just mixing them into a single number.
  • Increases confidence in laboratory test results, especially when results are used for diagnosis or research.

Disadvantages

  • Requires detailed planning and documentation — especially for reproducibility, which must be deliberately designed to include variations in days/operators/instruments, rather than just repeating measurements at random.
  • The time and costs required are greater if performed for many instruments or operators.
  • Interpretation can be complex when there is high variability that is not easily controlled — especially when having to statistically separate how much variation comes from within-day vs. between-day.

Data Required

  • Several identical samples tested repeatedly.
  • Test result data from repetitions under fixed conditions (repeatability).
  • Test result data from repetitions under different conditions (reproducibility) — for example, using different technicians, instruments, or times.
  • Typically processed using precision statistics, such as the coefficient of variation (CV%), standard deviation, or ANOVA (Analysis of Variance) to formally separate those sources of variation.

How to Use

Repeatability:

  1. Select 1 test sample.
  2. Perform the test at least 5–10 times in one session by the same person and instrument.
  3. Calculate the variation of the results (SD, CV%).

Reproducibility:

  1. Use the same test sample.
  2. Perform the test at different times, by different technicians, and/or with different instruments.
  3. Compare the results and analyze their variation.
In practice, the most common and practical design for laboratories is a simple nested design: several replicates measured every day (capturing repeatability), performed over several days in a row (capturing between-day reproducibility). From this design, ANOVA is used to mathematically separate the two sources of variation — explained in more detail in the next section.

Try the Precision Study Calculator

📐 Precision Study Calculator

Separate Repeatability (within-day) from Within-Laboratory Precision (within + between-day) using a nested ANOVA design.

Precision terminology

In ISO 5725, 'reproducibility' means between-laboratory. For a single lab, the standard terms are Repeatability (within-day) and Within-Laboratory Precision (within + between-day). This tool follows CLSI EP05-A3 / ISO 5725-3 (intermediate precision).

Quick Preset:

One row = one day. Separate replicates with commas.

Replicate count must be equal each day. Example: 20 days × 2 replicates (EP05-A3 lite) or 5 days × 5 replicates (EP15-A3).

Compare within-lab CV to a target imprecision. Leave empty to skip.

Repeatability (within-day)
SD0.443
CV%0.44%
df15
95% CI SD0.327–0.687
Between-day Component
SD0.987
df14
Within-Lab Precision (total)
SD1.082
CV%1.07%
df273
95% CI SD0.998–1.181

Between-day F-test

F = 10.915, p = 0.0000. Significant: real between-day variation exists.

ANOVA Table

Source of VariationSSdfMSFp
Between-day30.0020142.143010.9150.0000
Within-day (Repeatability)2.9450150.1963——
Total32.947029

Daily Means

13579111315Day9999.9100.8101.7102.6

Dots = daily means; green line = grand mean, band = ±2 SD within-lab.

Variance Decomposition

Within-day 16.8%Between-day 83.2%

Contribution of within-day vs between-day variance to total precision.

Daily Summary

DayReplicatesMean
1100.20, 101.10100.650
299.80, 100.50100.150
3101.50, 102.00101.750
498.90, 99.6099.250
5100.80, 101.30101.050
6102.10, 102.70102.400
799.50, 100.1099.800
8101.00, 101.60101.300
9100.30, 100.90100.600
1098.70, 99.3099.000
11101.80, 102.40102.100
12100.00, 100.60100.300
1399.20, 99.8099.500
14101.30, 101.90101.600
15100.60, 101.20100.900

Two-level nested design (day × replicate) with variance component (MSb−MSw)/n. Repeatability = √MS_within; Within-Lab = √(s_r² + s_day²). 95% CI for SD via chi-square (df = k(n−1) for repeatability; Satterthwaite for within-lab). Replicate definition: repeat of the same control material, same run, no recalibration — if the two values/day come from two different runs, what is computed is not pure repeatability. References: CLSI EP05-A3, CLSI EP15-A3, ISO 5725-3.

This calculator applies the nested design above: you input data grouped by day (with several replicates each day), and the calculator automatically performs the ANOVA calculation to separate Repeatability from Reproducibility.

The Logic Behind It: Simple Nested ANOVA

To avoid it being a "black box," here is an overview of the logic behind the calculator's calculation:

  1. Calculate the variation within each day (within-day) — how different the replicates are within the same day. This purely reflects repeatability, because measurement conditions within a single day are relatively uniform.
  2. Calculate the variation between daily means (between-day) — how different the result averages are from one day to the next. If the instrument is truly stable from day to day, the daily averages should be similar to one another.
  3. Combine the two into reproducibility — because in real daily conditions, the total variation experienced by the laboratory is a combination of both within-day variation and between-day variation.

Mathematically, this is why the Reproducibility SD is always greater than or equal to the Repeatability SD — reproducibility contains all the sources of variation present in repeatability, plus extra variation from day-to-day differences. A small Repeatability SD but a much larger Reproducibility SD is an important signal: the instrument is very consistent within a single work session, but something is shifting significantly from day to day — perhaps unstable daily calibration, the influence of changing room temperatures, or variations in technique between operator shifts.

Suitable for Use When

  • Validating or verifying a new method.
  • Replacing an instrument, technician, or reagent in the laboratory system.
  • As part of an internal quality audit or laboratory accreditation.
  • Evaluating EQC (External Quality Control) results or proficiency testing programs.

Conclusion

Repeatability and reproducibility are two crucial aspects of ensuring laboratory result reliability. Repeatability measures result stability under uniform conditions, while reproducibility tests whether results remain consistent when conditions change. Both provide a comprehensive picture of the precision of the methods and instruments used in the laboratory — and most importantly, separating them statistically provides much richer diagnostic information than just calculating a single combined CV number.

The implementation of these two instruments is essential in the context of clinical laboratory quality management. Through statistical analysis such as nested ANOVA — rather than just a single standard deviation — a laboratory can evaluate whether result variation is still within acceptable limits, and identify which source of variation needs the most improvement. This ensures that the test results provided to the clinician have a high level of confidence, regardless of who performed the test or when it was done.

Ready to streamline your QC workflow?

Track control values, get Levey-Jennings charts, and detect Westgard rule violations automatically.

Start Free at myqclab.com